Fluorescent X-ray film thickness meter

Last updated on June 28, 2024.

This project has been completed

The Yokohama Industrial Technology Support Center was abolished on March 31, 2024. Thank you very much for your continued patronage. Please note that the following is information from the past.

Introducing equipment for fluorescence X-ray film thickness meters

Appearance of fluorescent X-ray thickness meter


Overview

Principle

When X-rays drawn from the W target are applied to a metal thin film sample (such as a plating film), fluorescence X-rays are released according to the elements that make up the sample. The strength is measured separately for each energy (detector: proportional count tube, PIN detector). Since there is a correlation between film thickness and fluorescent X-ray strength, the film thickness can be determined by the obtained fluorescence X-ray strength.

Features

It is possible to know the thickness of the metal thin film (single-layer film, two-layer film, alloy film) on the material (metal, plastic). Since the sample is only irradiated with X-rays, it is in principle non-destructive measurement (however, depending on the size of the sample, it may not be possible to measure it without destructiveness). The measurable element range is from Ti (titanium) No. 22 to U (uranium) No. 92.

What I'm not good at

Fluorescent X-rays from the measuring part must be detected, so if there is a shield between the detector and the measuring point, it cannot be measured (for example, the thickness of the inner part of a concave sample).

In combinations containing the same metal (e.g. copper plating film on brass, etc.), it is impossible to separate the obtained fluorescence X-rays, so the membrane thickness cannot be measured.

Equipment Performance and Specifications

Device model number: FT9400 manufactured by Hitachi High-Tech Science Co., Ltd.
Date of introduction: November, 2014

Fees

Thickness test Fluorescent X-ray film thickness meter 2,400 yen per measuring point per sample.

Fees

  • Those that require special materials, labor, etc. for testing, analysis or preparation, and the amount of research or investigation fees shall be equivalent to actual costs.
  • In particular, although the deadline is set and urgent, the amount of the fee or fee for use shall be twice the amount specified.
  • Person who has office or establishment in Yokohama-shi, and the amount of fee or the fee for use pertaining to request from person except small and medium-sized business (outside site) prescribed in Article 2 of Small and Medium Enterprise Basic Law is 1.3 times the amount (If there is a fraction less than 100 yen, the fractional amount is rounded up to 100 yen.) It's a coincidence.
  • Fees or royalties pertaining to requests from persons who do not have Address in Yokohama City or individuals who do not have offices or offices in the city, corporations or other organizations are 1.5 times the amount specified (if there is a fraction less than 100 yen, the fractional amount is rounded up to 100 yen.) It's a coincidence.

Inquiries to this page

Economic Affairs Bureau Small and Medium Enterprise Fisheries Promotion Department Small and Medium Enterprise Promotion Division

Phone: 045-671-4236

Phone: 045-671-4236

Fax: 045-664-4867

E-Mail address [email protected]

Return to the previous page

Page ID: 997-983-840

Menu

  • LINE
  • Twitter
  • Facebook
  • Instagram
  • YouTube
  • SmartNews